V9400 Capnograph Technical Specification |
|
Parameters Monitored
CO2 Technique: Range:
Accuracy: Rise Time: Respiration Rate Range: Accuracy: Anaesthetic Agents Technique: Range:
Accuracy: Alarm Limits Ranges ETCO2 High:
Resp. Rate:
Serial Output:
Weight: Environment Temperature:
|
End Tidal CO2, Inspired CO2, Respiration Rate Anesthetic Agents Halothane,Isoflurane and Sevoflurane,
Non dispersive Infrared Spectroscopy 0-100mmHg, 0-13.3kPa, 0-10% +2 mmHg or 4% of reading, whichever is greater, 0-10% CO2, from 0-36 breaths/min 400ms (150ml/min flow)
+ 1 breath/min (from 0-120 breath/min)
Piezoelectric Crystal Resonance 0-8% Isoflurane, Halothane and Sevoflurane The greater of + 10% of reading or +0.2% from 0-6% accuracy outside of range is unspecified High/Low calibration required every 100 hours of operation
0-100 mmHg
(1mmHg steps) and OFF
High: 5-150 bpm
(1bpm steps) and OFF
High: 0.0-8.0%
(0.1 steps) and OFF
RS232C Data
Format 9600 baud, 1 start bit, 8 data bits, 1 stop bit no parity AC Supply Input of 200~240 VAC, 50/60Hz Output of 24 VDC @ 1.3A with 4kV isolation Dimensions Width:10.0 (2.4 cm) Height: 5 (8.9 cm) Depth: 5.5 (14.0 cm) Approximately 2.2 Kg (5 pounds)
Anesthetic
Agents Operating 15-30º C. Other Parameters Operating 0-50º C
Operating:
15-95% (non-condensing) 600mmHg to 760mmHg (0 to ~6,000 ft altitude)
|