burtons logo

picture of v9004

 

V9004 Capnograph

Technical Specifications

Displays & Indicators
Display

 
Vacuum Fluorescent

Pulse Tone

Pitch corresponds to SPO2 value.Volume adjustable

Indicators

Alarm, Alert, charge & silenced

FiO2(optional)
Range

 
0-100%

Displayed Trends

Graphical display, user selectable

Accuracy

± 2%

Displayed

15hr display @ 12sec. sample rate

Response Time

<10 sec to 90%

Waveforms

9hr display @ 4 sec.sample rate
Selectable CO2 or CO2/plethysmogram

Drift

<2% full scale per month

Alarms
Physiological

 
Dual pitch, adjustable volume.
User selectable high & low limits for all displayed parameters. Screen indicators for each parameter

Dimensions

 

Weight

3.5"x10.0"x5.5" HWD
88.9x254x139.7mm

 
5lbs; 2.27kg

System

Single pitch, adjustable volume

Power requirements

Silence

Temporary 2 mins. or OFF LED indicator

AC charger

230V, 50/60Hz

Flow Rate

150ml/min

Battery

Nickel Cadmium 6VDC

CO2
Measurement

 
Non-Dispersive IR Absorption

Internal, rechargeable 2 hrs usage, 4hrs Max charge time

Range

0-10% CO2

Serial Output

RS232C tabular trends,

Units

User Selectable: mmHg,Vol %,kPa

Patient data

Display Range

0-100mmHg, or 0-10.0% CO2 or
0-13.3kPa

Analog Output
Any 3 of CO2 waveform, Plethysmogram, ETCO2, Respiration rate, Inspired CO2, %SPO2, Pulse rate, Fractional inspired O2, 0V or 1V cal signal,
Range 0-1 VDC full scale.

Accuracy

±2mmHg; ±0.3%; ±0.3kPa, or 4% whichever is greater (0-10% CO2)

Resolution

1mmHg; 0.1%; 0.1kPA

Enviromental Specifications

Calibration

2 point Cal

Operation Temp.

0-50°C

N2O Compensation

User selectable 40% (default = OFF)

Storage Temp.

-40-70°C

Response Time

375msec to 90% of value

Relative Humidity

10-95% (storage)

Averaging

4 breath average

(non condensing)

15-95% (operating)

Respiration Rate
Range

 
0-150bpm

Safety Approvals
Designed to meet IEC601-1, CSA 125,UL544,IEC601-1-2 (EMC CE)

ISO 9001 Certified

Averaging

4 breath average

Accuracy

±1bpm

SpO2 (optional)
Range

 
0-100%

Accuracy

±2%@ 70-100%;±3%@50-69%

Averaging

4, 8 or 16 beat, user selectable

Pulse rate (optional)
Range

 
20-350bpm

Accuracy

± 1 beat or 2% whichever is greater

Averaging

8 or 16 seconds, user selectable

Link to patient monitors

Use this link to download a brochure for the V3402

Specifications subject to change without notice