Burtons Logo

Picture of 9401

 

9401 Non Invasive Blood Pressure Monitor

Technical Specification

Technique:
Parameters:
Modes:
History Mode:
Automatic Cycle Times:
Determination Time:
 
Initial Inflation:
Pulse Rate Range:
Maximum time allowed:
Blood Pressure Range:
BP Accuracy:
Pulse Accuracy:
Patient Alarms:


Power

 
Source:
AC Power:
Battery:
Leakage current:

Printer:
PC Connection:
Operating Environment
Temperature:
Humidity:
Altitude:
Physical
Dimensions: Weight:

Oscillometric measurement method.
Systolic, diastolic, mean arterial pressure and heart rate (20-300 beatsper minute).
Manual, Auto Cycle, STAT (5 min. consec.)
Maintains last 5 hours up to 99 readings.
1, 2, 3, 4, 5, 10, 15, 30, 60, 90.
Small cuff: 18 seconds typical @160 BPM 120/80 
Large cuff: 26 seconds typical @75 BPM 120/80
150 mmHg; subsequent inflations to approx. 30 mmHg greater than previous systolic pressure.
20-300 BPM
150 seconds.
20-265 mmHg
 ±5 mmHg with standard deviation no greater than 8 mmHg. Meets AAMI/ANSI SP10, '92
 ±2%
  High Limit Low Limit Preset
Systolic 80-250 40-160 H:240/L:0
Diastolic 30-200 30-150 H:130/L:0
MAP 50-220 35-150 H:220/L:0
Pulse Rate 20-300 20-300 H:290/ L:0
AC line or internal battery.
90-264VAC, 50-60 Hz, 0.3A.
Nickel metal hydride battery pack; charge time is less than 4 hours (easily changed for extended battery life); operation on battery is 100 readings @ 1 per minute.
100 micro amp maximum
(Optional) RS232 port for printer connection.
Use NULL Modem cable to RS232 port.
 0°C to 50°C (32°F to 122°F).
15-95%, non-condensing.
500 feet (152 meters) below sea level to 10,000 feet (3048 meters) above sea level.
Storage Temperature: -20°C to 70°C (-40°F to 158°F).
17.1cm H x 21.6cm W x 7.6cm D / 6.75" H x 8.5"W x 3.0"D
1.4 kg (3 Ibs.)

Specifications Subject to Change Without Notice